
80
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
This window also allows you to enter some general information about the transmitter, such as
Message, Tag, Descriptor, and Date. This information is not required for the operation of the
transmitter and can be left out if desired.
4.6.8 Setup - Output
The Output tab lets you assign up to four transmitter variables.
Typically, the PV is configured to be Product Level, Interface Level, or Volume.
Other variables like Product Distance, Interface Distance, Upper Product Thickness, etc. are
available as well.
For Rosemount 3301, the PV is typically set to be Level. If the transmitter is in the Immersed
Probe mode (see Measurement mode on page 71), the PV is normally set to Interface Level.
For Rosemount 3302, the PV is typically set to be Interface Level, but Level and other options can
also be used.
Set the Lower Range Value (4 mA) and the Upper Range Value (20 mA) to the desired values.
Keep in mind that the 20 mA value must be below the upper transition zone, and the 4 mA point
must be above the lower transition zone if you want to use the full 4-20 mA range within the
measuring range of the transmitter.
Make sure the 20 mA value is set below the Upper Null Zone (UNZ). (The UNZ parameter can be
used if there are measurement problems in the upper part of the tank, see Disturbances at the
top of the tank on page 102.) In the default configuration, the UNZ is equal to zero.
See Transition zones on page 16 for more information on upper and lower transition zones.
See Basic configuration on page 69 for more information on setting the Upper and Lower Range
values.
The default Damping value is 10. Normally this value does not need to be changed. The
Damping parameter may be changed if there are high filling rates, see High level rates on
page 98 for more information.
Kommentare zu diesen Handbüchern