27
Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
For information on the recommended minimum free space (L) to tank wall or other objects in
the tank, see Table 3-5.
Table 3-5. Recommended Minimum Free Space (L)
For information on the recommended minimum free space (L) to tank wall or other objects in
the tank in the case of rigid single, segmented rigid single and flexible single lead probes, see
Table 3-6.
Table 3-6. Free Space (L) Requirements - Single Lead Probes
3.4.5 Recommended mounting position
When finding an appropriate mounting position for the transmitter, the conditions of the tank
must be carefully considered. The transmitter should be mounted so that the influence of
disturbing objects is reduced to a minimum.
In case of turbulence, the probe may need to be anchored to the bottom. See “Mechanical
installation” on page 29 for more information.
Figure 3-9. Mounting Position
Coaxial Rigid twin Flexible twin
0 in. (0 mm) 4 in. (100 mm) 4 in. (100 mm)
Rigid single/segmented rigid single/flexible single
4 in. (100 mm) Smooth metal wall
12 in. (300 mm)
Disturbing objects, such as pipes and beams, concrete or plastic tank walls,
rugged metal tank walls
A
C
B
A Inlet pipe
B Agitator
C Heating coils
Kommentare zu diesen Handbüchern